Instructions For Use
INSTRUCTIONS FOR USE
Foreign Body Removal Kit
applicable to Standard, Deluxe, and custom kit assemblies
| Document Number | IFU-FBR-001 |
| Effective Date | 05/08/2026 |
| Manufacturer | Medical Device Purchase, LLC |
| Customer Service | [email protected] |
| Web | buymdp.com |
1. Product Description
The Foreign Body Removal Kit is a pre-assembled set of stainless-steel ophthalmic instruments and accessories used by qualified eyecare professionals for the examination and removal of superficial foreign bodies and rust from the cornea and adjacent ocular tissues. The kit is supplied non-sterile and is intended for repeated use after appropriate cleaning and sterilization.
Typical Kit Contents
Specific kit contents vary by configuration (Standard, Deluxe, or custom). A typical Foreign Body Removal Kit includes the following components:
- Algerbrush II handpiece with 5 mm or 1.0 mm burr (battery-operated rotary instrument for rust-ring removal)
- Replacement burr tip(s)
- Foreign body spud or golf-club spud (single-ended or double-ended)
- Jeweler’s forceps, fine-tip
- Eyelid speculum (wire or solid blade)
- Sterilization tray with silicone mat or instrument holder
- Instructions for Use (this document)
Refer to the kit’s outer label and packing list for the exact contents of your specific configuration.
Component model numbers, sizes, and quantities are listed at buymdp.com.
Materials
- Instruments: Surgical-grade stainless steel (AISI 420 / AISI 304 or equivalent).
- Sterilization tray: Stainless steel with silicone
- Algerbrush II handpiece: Anodized aluminum body with stainless-steel burr; powered by one (1) AA alkaline battery (not included).
Classification
- FDA Device Class: Class I (reusable surgical instrument; some configurations may include Class II components).
- Sterility State as Supplied: Non-sterile. Must be cleaned and sterilized prior to first use and between uses.
- Single Use / Reusable: Reusable, except where individual components are explicitly labeled “Single Use Only.”
2. Intended Use, Indications, and Users
Intended Use
The Foreign Body Removal Kit is intended for the examination, manipulation, and removal of superficial foreign bodies (e.g., metallic particles, rust, organic debris) from the cornea, conjunctiva, and adjacent ocular tissues in clinical and surgical settings.
Indications for Use
- Removal of superficial corneal or conjunctival foreign
- Removal of corneal rust rings following metallic foreign-body
- Atraumatic eyelid retraction during ophthalmic examination and minor
- Adjunctive use during ophthalmic examination, minor surgical procedures, and emergency or urgent-care evaluation of the eye.
Intended Users
This device is intended for use only by licensed eyecare professionals (ophthalmologists, optometrists, emergency-medicine physicians) and other qualified clinicians who have received appropriate training in ophthalmic instrument use and aseptic technique. The device is not intended for use by patients or untrained personnel.
Intended Patient Population
Adult and pediatric patients presenting with a superficial ocular foreign body or rust ring, where removal is clinically indicated. Use in pediatric patients should be at the discretion of the treating clinician with appropriate restraint and patient-cooperation considerations.
Intended Use Environment
- Clinical exam rooms, urgent-care facilities, and emergency
- Ambulatory surgery centers and ophthalmic operating
- Veterinary ophthalmology clinics (where applicable).
3. Contraindications, Warnings, and Precautions
Contraindications
- Do not use on patients with known hypersensitivity to any component
- Do not use for the removal of intraocular or deeply embedded foreign Such cases require referral to ophthalmology and may require operative intervention.
- Do not use in cases of suspected globe rupture, full-thickness corneal laceration, or open globe injury. Removal attempts are contraindicated and may worsen ocular injury.
- Do not use on patients with severe ocular infection where instrumentation may exacerbate the condition, except as clinically warranted.
WARNING: Do not attempt removal of foreign bodies that appear to penetrate the corneal stroma or anterior chamber. Refer to ophthalmology emergently. Inappropriate use of removal instruments on penetrating injuries may cause permanent vision loss.
Warnings
- Inspect every instrument before each Do not use any instrument that is bent, cracked, corroded, has a damaged tip, or shows signs of insulation damage (where applicable).
- Use only as directed and only by trained
- Do not exceed the maximum sterilization temperature specified in Section Exceeding recommended temperatures may damage instruments and compromise sterility.
- The Algerbrush II rotary burr can cause iatrogenic corneal injury if used Apply only light pressure and stop immediately if the patient moves or if the instrument stalls.
- Use appropriate personal protective equipment (PPE) including gloves and eye protection during use, cleaning, and sterilization.
CAUTION: Sharps and pointed instruments included in this kit may cause puncture injury. Handle with care, dispose of damaged instruments in an approved sharps container, and follow your facility’s sharps-handling protocol.
Precautions
- Confirm patient identity and obtain informed consent prior to any procedure, in accordance with your facility’s policies.
- Apply topical anesthetic per clinician judgment and standard of care prior to
- Use adequate magnification (slit lamp or operating microscope) for foreign-body removal procedures whenever possible.
- Do not modify the Modification voids all warranties and may render the device unsafe.
- Single-use components labeled as such must be discarded after use and must not be
Potential Adverse Effects
- Corneal abrasion, scarring, or
- Conjunctival hemorrhage or
- Infection (bacterial keratitis, conjunctivitis).
- Iatrogenic deepening of an existing corneal
- Allergic or hypersensitivity reaction to instrument materials (rare).
Report any serious incident that occurs in relation to this device to Medical Device Purchase, LLC and to the competent regulatory authority of the jurisdiction in which the user and/or patient is established.
4. Instructions for Use
4.1 Pre-Use Inspection
- Verify the kit is complete using the packing
- Confirm all instruments have been cleaned and sterilized in accordance with Sections 5 and
- Inspect each instrument under adequate light for damage, corrosion, burrs, or misalignment of
- For Algerbrush II: insert a fresh AA alkaline battery, attach the appropriate burr, and confirm smooth rotation. The burr should turn freely and stop immediately when light pressure is
- Set up the procedure tray with all required instruments within easy
4.2 Procedure
- Position the patient comfortably at a slit lamp or operating microscope with adequate
- Instill topical ophthalmic anesthetic per institutional
- Insert the eyelid speculum gently to maintain palpebral Avoid contact with the cornea.
- Identify and characterize the foreign body and the depth of
- For superficial foreign bodies: use the spud or jeweler’s forceps to lift the foreign body off the corneal or conjunctival surface using the smallest motion possible.
- For rust rings or embedded metallic debris: assemble the Algerbrush II and apply the rotating burr tangentially to the lesion using minimal pressure. Use brief, controlled passes. Stop frequently to assess depth and clearance.
- Irrigate the eye as needed with sterile
- Inspect the eye for residual material, depth of defect, and adjacent corneal
- Remove the speculum
- Apply post-procedure treatment (antibiotic, lubricant, patch, ) per clinician judgment and standard of care.
- Document the procedure, instruments used, and any complications in the patient
4.3 Post-Use Handling
- Remove single-use components and dispose of them as biohazardous
- Remove the burr from the Algerbrush II Discard used burrs as sharps.
- Remove the AA battery from the Algerbrush II handpiece if it will not be used Do not reprocess the handpiece with the battery installed.
- Pre-clean reusable instruments at the point of use by wiping visible debris and immersing instruments in distilled water or an enzymatic pre-soak solution to prevent drying of
- Transport instruments to the reprocessing area in a closed, leak-proof container labeled as
IMPORTANT: Do not allow blood, saline, or tissue debris to dry on instruments. Dried bioburden is significantly more difficult to remove and can compromise the effectiveness of subsequent cleaning and sterilization.
5. Cleaning and Disinfection
Reusable instruments must be cleaned thoroughly before sterilization. Cleaning effectiveness directly affects sterilization effectiveness.
5.1 Manual Cleaning
- Don appropriate PPE (gloves, gown, mask, eye protection).
- Disassemble all instruments to the extent possible (e.g., remove burrs, open hinges).
- Rinse instruments under cool running water to remove gross Do not use hot water, which can coagulate proteins.
- Prepare a neutral-pH (pH 0–8.0) enzymatic detergent solution per the detergent manufacturer’s instructions.
- Fully immerse instruments and brush all surfaces, hinges, serrations, and lumens with a soft- bristle brush. Pay particular attention to box locks, ratchets, and tips.
- Soak for the dwell time recommended by the detergent manufacturer (typically 5–10 minutes).
- Rinse thoroughly with deionized or distilled water for a minimum of 30 seconds, ensuring all detergent residue is removed.
- Dry instruments completely with a clean, lint-free cloth or filtered medical-grade compressed
5.2 Ultrasonic Cleaning (Recommended)
- Place instruments in the ultrasonic bath in an open position; do not stack or
- Use a neutral-pH enzymatic detergent at the concentration and temperature recommended by the detergent manufacturer.
- Run the ultrasonic cycle for the manufacturer-recommended duration (typically 10–15 minutes at 35–45 °C).
- Rinse and dry as described in Section 1.
5.3 Algerbrush II Handpiece
- Do not immerse the Algerbrush II handpiece in liquid or place in an ultrasonic
- Wipe the exterior of the handpiece with a soft cloth lightly moistened with a neutral-pH detergent solution, followed by a cloth moistened with distilled water.
- Allow the handpiece to air-dry completely before storage or sterilization (where the handpiece manufacturer’s IFU permits).
- Refer to the Algerbrush II manufacturer’s separate Instructions for Use for full reprocessing requirements.
5.4 Inspection After Cleaning
- Inspect each instrument under magnification for visible soil, corrosion, pitting, or
- Verify that hinges, ratchets, and moving parts operate
- Re-clean any instrument that is not visibly Remove from service any instrument that is damaged.
6. Sterilization
Sterilize all reusable instruments in the kit prior to first use and after every subsequent use. Validated sterilization parameters for stainless-steel ophthalmic instruments are summarized below.
6.1 Steam Sterilization (Autoclave)
| Cycle Type | Temperature | Minimum Exposure Time |
| Pre-vacuum (dynamic-air- removal) | 132 °C / 270 °F | 4 minutes |
| Pre-vacuum (dynamic-air- removal) | 134 °C / 273 °F | 3 minutes |
| Gravity displacement | 121 °C / 250 °F | 30 minutes |
| Gravity displacement | 132 °C / 270 °F | 15 minutes |
- Minimum dry time: 20–30 minutes (or per autoclave manufacturer recommendation).
- Use FDA-cleared sterilization wraps or rigid sterilization containers compatible with the selected
- Do not exceed 134 °C / 273 °F. Higher temperatures may damage instruments and shorten service life.
WARNING: Flash sterilization (immediate-use steam sterilization) is not recommended for routine reprocessing of these instruments. Flash sterilization should be limited to documented emergency use only and in accordance with AAMI ST79 and your facility’s policies.
Allow the load to cool to room temperature before handling. Do not force-cool with cold water, which may warp instruments.
6.2 Chemical Sterilization
Chemical sterilants compatible with stainless-steel ophthalmic instruments may be used in accordance
with the sterilant manufacturer’s validated instructions. Acceptable chemistries include:
- Hydrogen peroxide gas plasma (e.g., STERRAD®) per system manufacturer’s validated
- Ortho-phthalaldehyde (OPA) or glutaraldehyde for high-level disinfection only — not equivalent to sterilization.
- Ethylene oxide (EtO) per ANSI/AAMI/ISO
Always verify chemical compatibility with all instrument materials prior to use. Rinse thoroughly with sterile water following chemical exposure to remove residues.
6.3 Sterilization Tray
- Verify that the sterilization tray is rated for the selected cycle and load
- Distribute instruments evenly across the Do not overload.
- Ensure that hinged instruments are placed in the open position to allow steam
6.4 Sterility Assurance
- Use chemical and biological indicators in every load in accordance with AAMI
- Document each cycle (load contents, parameters, indicator results, operator) and retain records per institutional policy.
- Reprocess any load that fails biological or chemical indicator
7. Storage, Shelf Life, and Service Life
Storage
- Store sterilized instruments in a clean, dry, environmentally controlled area, protected from dust, moisture, and direct sunlight.
- Recommended storage conditions: 15–25 °C, relative humidity 30–70%, no exposure to corrosive vapors.
- Maintain sterile barrier Reprocess any pack with a compromised sterile barrier (tear, moisture, broken seal).
Shelf Life of Sterile Pack
Shelf life of a sterile pack is event-related, not time-related. A pack remains sterile until a sterility- compromising event occurs (drop, tear, moisture exposure, broken seal). Follow your facility’s event- related sterility maintenance policy.
Service Life of Instruments
When properly cleaned, sterilized, stored, and inspected, the reusable instruments in this kit are designed for repeated use over a service life consistent with industry standards for surgical-grade stainless-steel instruments. Replace any instrument that shows wear, corrosion, dull or damaged tips, misalignment, or compromised function.
8. Symbols Glossary
The following symbols may appear on device labeling, in accordance with ISO 15223-1 and 21 CFR 801.15:
| Symbol | Meaning | Standard |
| MD | Medical Device | ISO 15223-1 |
| REF | Catalogue Number | ISO 15223-1 |
| LOT | Batch / Lot Number | ISO 15223-1 |
| NON-STERILE | Non-sterile as supplied | ISO 15223-1 |
| Rx Only | Caution: U.S. Federal law restricts this device to sale by or on the order of a licensed practitioner. | 21 CFR 801.109 |
| Manufacturer | Indicates the manufacturer | ISO 15223-1 |
| Date of Manufacture | Indicates the date the device was manufactured | ISO 15223-1 |
| Consult IFU | Refer to instructions for use / electronic instructions for use | ISO 15223-1 |
| Single Use | Do not reuse | ISO 15223-1 |
9. Warranty
Medical Device Purchase, LLC warrants the reusable stainless-steel instruments in this kit against defects in materials and workmanship for the lifetime of the instrument under normal use. The warranty also covers corrosion under normal use when the instruments have been cleaned, sterilized, and stored in accordance with this IFU.
This warranty does not cover: damage caused by misuse, modification, exposure to inappropriate chemicals or sterilization parameters; normal wear of cutting edges, tips, or working surfaces; battery- operated components beyond their stated service life; or single-use items.
Warranty claims should be submitted to [warranty email] with proof of purchase, photographs of the affected instrument, and a description of the issue. See buymdp.com for the full warranty policy.
10. Disposal
- Dispose of instruments at end of service life in accordance with applicable local, state, and federal regulations governing medical-device disposal and metal recycling.
- Dispose of contaminated single-use components and used burrs as biohazardous sharps waste
in accordance with your facility’s sharps-handling protocol.
- Remove batteries from the Algerbrush II handpiece prior to disposal and dispose of batteries in accordance with applicable battery-recycling regulations.
11. Reporting Adverse Events
Healthcare providers should report any serious adverse event associated with the use of this device to:
- Medical Device Purchase, LLC – [email protected]
- S. Food and Drug Administration MedWatch program — www.fda.gov/medwatch (1-800-FDA- 1088).
- Competent regulatory authority of the user’s and/or patient’s jurisdiction (for users outside the United States)